TGS autoclaves include 5 fixed sterilization programs, 2 test programs and 22 customizable sterilization programs.
Program
Load Type
Temp.
(oC)
Sterilization
Time (min)
Dry Time
(min)
Total Time without drying (min) *
5075 GS
50 125 GS
Unwrapped 134
Unwrapped instruments
134
4
1
24
21
Wrapped 134
Wrapped instruments
134
7
20
37
32
Unwrapped 121
Unwrapped delicate instruments
121
20
1
36
34
Wrapped 121
Wrapped delicate instruments
121
20
20
50
41
Prion 134
Prion
134
18
20
50
47
Leak Test
No load
-
-
-
31
20
Bowie & Dick Test
PCD
134
3.5
2
21
25
* Cycle times are based on: 15kg load for 5075GS, 33 kg load for 50125GS, and do not include drying. Cycle times based on warm chamber and may change according to load size. Bowie & Dick Test is also suitable for Helix test device. Leak Test checks that the chamber under vacuum does not leak.