Cycle Programs

5 pre-set programs, 2 test programs Bowie & Dick and Leak test, and 20 customizable programs.
Bowie & Dick test, also suitable for Helix test device, includes 3.5 min sterilization time without drying.
Leak test checks that the chamber under vacuum does not leak.

 

Cycle Program
Load Type
Temp.
(oC)
Sterilization
Time (min)
Dry Time
 (min)
Cycle Time
 (min) *
Unwrapped 134 Unwrapped instruments 134 4 1 17
Wrapped 134
Wrapped instruments
134
4
20 24
Unwrapped 121
Unwrapped delicate instruments
121
20
1 32
Wrapped 121
Wrapped delicate instruments
121
20
20 34
Prion 134
Prion
134
18
20 34

 

* Cycle times are based on 1.5 kg standard load and do not include drying. Cycle times based on warm chamber and may change according to load size.