Pre-Programmed Sterilization Cycle Programs

The 3870 HSG autoclave includes 5 fixed sterilization programs, 2 test programs and 22 customizable sterilization programs.

 

Program
Load Type
Temp.
(oC)
Sterilization
Time (min)
Dry Time
 (min)
Total Time without drying
 (min) *
Unwrapped 134 Unwrapped instruments 134 4 1 31
Wrapped 134
Wrapped instruments
134
7
20 34
Unwrapped 121
Unwrapped delicate instruments
121
20
0 33
Wrapped 121
Wrapped delicate instruments
121
20
20 46
Prion 134
Prion
134
18
20 45
Leak Test
No load
-
-
- 24
Bowie & Dick Test PCD 134 3.5 2 12

 

* 3870 HSG cycle times are based on a 5kg load and  do not include drying. Cycle times based on warm chamber and may change according to load size.   Bowie & Dick Test is also suitable for Helix test device. Leak Test checks that the chamber under vacuum does not leak.