Place unit on a sturdy level counter.
-
Fill the Reservoir with distilled water
-
The unit is ready to operate
To Adjust the Filling Time:
-
Press the STOP Key repeatedly until the message “Code:xxx” appears.
-
Using the UP/DN arrow keys change the code to 105, then press the STOP Key.
-
A message will be displayed saying “Water in = xx sec”
-
Using the UP/DN arrow keys change the seconds according to the following:
EZ9 = 30 sec
EZ10 = 35 sec
3870EK = 65 sec
Now press the STOP Key
-
On some older units the message “Ea Type:” will appear,using the UP/DN arrow keys select either “0” for an E or EK type unit or “1” for an EA or EKA type unit.
-
Press the STOP Key to finish
-
The unit is ready to operate
For units with a Microprocessor version # NOT ending in WP
-
Follow steps 1–5 for Manual Units above,then empty the water from the chamber and check the automatic water fill.
-
Turn the unit on
-
With the door open press and hold the door switch,then press the start button. When water starts flowing release the door switch.
-
Water should come up to the same spot as the measured amount had.
-
If need be adjust the automatic fill using the following procedure.
To Adjust the Automatic Filling:
-
While leaving the door open turn power off.
-
Press and hold the Manual Water Fill Button (this is the button with the two arrows).
-
Turn the power back on
-
When the normal display screen appears release the fill button –immediately press it in again and monitor the water flow
-
Continue to hold the button until water reaches groove at front –then release button – wait ten seconds – the unit is reprogrammed.